1. Original imported high frequency host and high speed tube
-TOSHIBA highest inverter frequency x-ray tube and high frequency voltage generator.
Constant direct current high voltage output, it can achieve high quality monochromatic X-Ray and eliminate the harmful effect of soft ray to image thoroughly.
-Equipped with high performance and large capacity X-Ray tube
Adopt 0.6/1.2mm2 dual focus, 300KHU large capacity and high speed X-Ray tube,it is suitable for long time and high intensity clinical examination.
-Intelligent high voltage control system with multiple human APR, LCD display show APR rules, and set according patients figure, position and part automatically, it’s fast and convenient.
-Computer progress control, remote operate, electric multi-lead collimator, adjust X-ray range as diagnosis need, this reduce X-ray harm for patients and atmosphere
-Multi self-diagnosis program, has RS232 port, and with error indication function.
2. High definition digital image chain to ensure that minor lesions can not be passed.
-Imported DR flat panel detector, it does not need any conversion but it can finish digital photography of each position to achieve real direct digital.
-DR flat panel detector DR connected with image processor achieves low noise and rich contrast image, edge enhancement filtering device makes the edge of image more clear and sharp.
-3k×3k gathering matrix, more than 900 mega pixel, promise high photo space definition, and good signal-no-noise ratio.
-17”*17” fixed flat panel detector can cover whole abdomen and whole chest, no limitation for photography.
-Fast image speed, work together with DROC soft, and promote work efficiency.
-A-si stable function, low requirement for temperature, and no crystal under high temperature.
3. Equipped with new digital image system achieves rich digital image processing function
Advanced digital image processing function
-Pre-assembled Windows professional operating system and professional image processing software
Image playback: thumbnails view, sequence replay tools.
Image processing: W/L adjustment, arrows and words etc labels, angle and distance measurement, image scaling, translation, up/down conversion, left/right conversion, rotation, black and white reversal.
-Image storage: image real-time storage, DICOM image send, A shampoo Burning Studio, derived storage (choose various storage ways Bitmap, JPEG, AVI etc to be used in Word and Power-point office software, it is convenient for the doctors to write diagnostic reports and papers ).
-DICOM3.0: can be connected with laser camera to print film and PACS NETWORK.
-Medical record management: database management, graphic report, support for WORKLIST.
4. DR Workstation
Touch screen operation and UPS for the workstation
Computer system:
B150 PLUS motherboard; processor: Intel Core I5 6400
Memory: 8 GB DDR3
Hard disk: 1T SATA (7200 rpm)
Optical drive: DVR-221CHV CD burner
NIC: Intel(R)gigabit CT
Graphics card: gtx1050
Display: Dell (DELL) U2412Mc display
Main Technical Parameters:
Item |
Content |
Technical Parameters |
Power |
Voltage |
380V±10% |
Frequency |
50Hz±1Hz |
|
Capacity |
≥65kVA |
|
Internal Resistance |
≤0.17Ω |
|
High Frequency High Voltage X-Ray Generator
|
Power |
56KW |
Inverter frequency |
30 KHz |
|
Tube voltage |
40kV—150kV ,continuous adjustment |
|
Tube current |
10mA—630mA |
|
Exposure time |
01-6300ms |
|
mAs |
0.1-900mAs |
|
X-ray source assembly (TOSHIBA E7252X) |
Tube Focus:big/small |
1.2mm /0.6mm |
Input power |
Big focus 75kW Small focus 27kW |
|
Anode thermal capacity |
210KJ |
|
Rotary anode speed |
2800rpm |
|
Digital Flat Panel Detector |
Active area |
427(H)×427(V) |
Pixel matrix: |
3072(H)×3072(V) |
|
Pixel pitch: |
139 μm |
|
Cycle time: |
Less than 6 s |
|
Limiting resolution |
3.1 lp/mm typ |
|
A / D transition: |
14 bit |
|
Energy range |
40 - 150 kVp |
|
Maximum entrance dose (low gain) |
4 mR / frame |
|
power Input: |
DC 24V2A |
|
Bucky Stand |
Up and down movement |
Not less than 1300mm |
Table |
SID |
500~1280mm |
Table bearing |
135kg |