1. Original imported high frequency host and high speed tube
-TOSHIBA highest inverter frequency x-ray tube and high frequency voltage generator.
Constant direct current high voltage output, it can achieve high quality monochromatic X-Ray and eliminate the harmful effect of soft ray to image thoroughly.
-Equipped with high performance and large capacity X-Ray tube
Adopt 0.6/1.2mm2 dual focus, 300KHU large capacity and high speed X-Ray tube,it is suitable for long time and high intensity clinical examination.
-Intelligent high voltage control system with multiple human APR, LCD display show APR rules, and set according patients figure, position and part automatically, it’s fast and convenient.
-Computer progress control, remote operate, electric multi-lead collimator, adjust X-ray range as diagnosis need, this reduce X-ray harm for patients and atmosphere
-Multi self-diagnosis program, has RS232 port, and with error indication function.
2. High definition digital image chain to ensure that minor lesions can not be passed.
-Imported DR flat panel detector, it does not need any conversion but it can finish digital photography of each position to achieve real direct digital.
-DR flat panel detector DR connected with image processor achieves low noise and rich contrast image, edge enhancement filtering device makes the edge of image more clear and sharp.
-3k×3k gathering matrix, more than 900 mega pixel, promise high photo space definition, and good signal-no-noise ratio.
-17”*17” fixed flat panel detector can cover whole abdomen and whole chest, no limitation for photography.
-Fast image speed, work together with DROC soft, and promote work efficiency.
-A-si stable function, low requirement for temperature, and no crystal under high temperature.
3. Equipped with new digital image system achieves rich digital image processing function
Advanced digital image processing function
-Pre-assembled Windows professional operating system and professional image processing software
Image playback: thumbnails view, sequence replay tools.
Image processing: W/L adjustment, arrows and words etc labels, angle and distance measurement, image scaling, translation, up/down conversion, left/right conversion, rotation, black and white reversal.
-Image storage: image real-time storage, DICOM image send, A shampoo Burning Studio, derived storage (choose various storage ways Bitmap, JPEG, AVI etc to be used in Word and Power-point office software, it is convenient for the doctors to write diagnostic reports and papers ).
-DICOM3.0: can be connected with laser camera to print film and PACS NETWORK.
-Medical record management: database management, graphic report, support for WORKLIST.
4. DR Workstation
Touch screen operation and UPS for the workstation
Computer system:
B150 PLUS motherboard; processor: Intel Core I5 6400
Memory: 8 GB DDR3
Hard disk: 1T SATA (7200 rpm)
Optical drive: DVR-221CHV CD burner
NIC: Intel(R)gigabit CT
Graphics card: gtx1050
Display: Dell (DELL) U2412Mc display
Main Technical Parameters:
Item
|
Content
|
Technical Parameters
|
Power
|
Voltage
|
380V±10%
|
Frequency
|
50Hz±1Hz
|
Capacity
|
≥65kVA
|
Internal Resistance
|
≤0.17Ω
|
High Frequency High Voltage X-Ray Generator
|
Power
|
56KW
|
Inverter frequency
|
30 KHz
|
Tube voltage
|
40kV—150kV ,continuous adjustment
|
Tube current
|
10mA—630mA
|
Exposure time
|
01-6300ms
|
mAs
|
0.1-900mAs
|
X-ray source assembly
(TOSHIBA E7252X)
|
Tube Focus:big/small
|
1.2mm /0.6mm
|
Input power
|
Big focus 75kW Small focus 27kW
|
Anode thermal capacity
|
210KJ
|
Rotary anode speed
|
2800rpm
|
Digital Flat Panel Detector
|
Active area
|
427(H)×427(V)
|
Pixel matrix:
|
3072(H)×3072(V)
|
Pixel pitch:
|
139 μm
|
Cycle time:
|
Less than 6 s
|
Limiting resolution
|
3.1 lp/mm typ
|
A / D transition:
|
14 bit
|
Energy range
|
40 - 150 kVp
|
Maximum entrance dose (low gain)
|
4 mR / frame
|
power Input:
|
DC 24V2A
|
Bucky Stand
|
Up and down movement
|
Not less than 1300mm
|
Table
|
SID
|
500~1280mm
|
Table bearing
|
135kg
|